It is possible to pre-screen crystals mounted on I23 sample holders on the beamline I03. We recommend this procedure for crystals that show a high variability in diffraction quality. Since the screening time is much faster on I03 than on I23, it is a more efficient way to select the best crystals for anomalous data collection on I23.
To be able to compare the crystals diffraction quality for pre-screening, we recommend using the same data collection settings during the i03 beamtime. See below an example of screening settings. The exposure time might vary with your crystal system. Since the I23 beam typically operates with unfocussed X-ray beam, please select the largest beam on I03 (80x20 um).
If you have any question, please contact the I23/I03 staff or consult the beamline manuals.
https://www.diamond.ac.uk/Instruments/Mx/I23/I23-Manual.html
https://www.diamond.ac.uk/Instruments/Mx/I03/I03-Manual.html
Diamond Light Source is the UK's national synchrotron science facility, located at the Harwell Science and Innovation Campus in Oxfordshire.
Copyright © 2022 Diamond Light Source
Diamond Light Source Ltd
Diamond House
Harwell Science & Innovation Campus
Didcot
Oxfordshire
OX11 0DE
Diamond Light Source® and the Diamond logo are registered trademarks of Diamond Light Source Ltd
Registered in England and Wales at Diamond House, Harwell Science and Innovation Campus, Didcot, Oxfordshire, OX11 0DE, United Kingdom. Company number: 4375679. VAT number: 287 461 957. Economic Operators Registration and Identification (EORI) number: GB287461957003.