Beamline Phone Number:
+44 (0) 1235 778917
Principal Beamline Scientist:
Kawal Sawhney
Tel: +44 (0) 1235 778169
E-mail: [email protected]
Restrictions apply. Please contact beamline staff well in advance to discuss feasibility.
Feature | Availability | Comments |
Monochromatic focused (4-20 keV) | Available | Typical photon flux: ~1 x 1012 ph/sec |
Monochromatic unfocused (4-45 keV) | Available | Typical photon flux: ~1 x 109 ph/sec/mm2 |
White beam | Available | |
Monochromatic, focused or unfocused (2–4 keV) | Limited availability | Very low flux as beam exits through a Be window. Contact B16 staff well in advance. |
Beam size: Focused < 200 μm x 200μm Unfocused: 100mm (h) x10 mm(v) Micro-focused with CRLs : ~ 2 μm x 3 μm with KB mirrors: ~ 0.5 μm x 0.5 μm | Available | Typical photon flux in micro-focused beam: ~1 x 108 ph/sec |
Equipment | Availability | Comments |
Double Crystal Monochromator Si 111 (2 – 30 keV) | Available | |
Channel-cut Monochromator Si 311 (6 - 45 keV) | Limited availability | |
Double Multilayer Monochromator - for higher-order suppression and for high flux modes - Ni/B4C: 2- 8 keV - Ru/B4C: 8-20 keV | Available | |
Energy bandpass - DCM Si 111 : ~1.8 x10-4 - Channel-cut Si 311 : ~1 x 10-4 - Multilayer : ~1 x 10-2 | Available | |
Toroidal mirror (1:1 focusing) | Available | Focal distance of the mirror can be adjusted to keep the beam in-focus at any position along the 12m length of the experimental hutch. |
Micro focusing optics - Be compound refractive lenses
| Available | With CRLs, beam size is energy / focal length dependent. |
Equipment | Availability | Comments |
Large area XRD detector: Photonic Science X-ray Image Star 9000, 135mm dia, 3056 x 3056 pixels, 31 µm optical pixel size | Available | |
Pilatus 300k detector: 84mm x 106 mm, 487 x 619 pixels, 172 µm x 172 µm pixel size, 20 bit dynamic range | Available | |
Medipix Merlin 3.0 detector: 14 mm x 14 mm, 256 x 256 pixels, 55 µm pixel size | Available | |
Topography camera: Photonic Science X-ray sCMOS topography camera, 7mm dia, 1920 x 1080 pixels, ~3.2μm optical pixel size, 14 bit dynamic range | Available | |
VORTEX 90EX spectroscopy detector: Active area = 50mm2 Energy resolution ~ 150 eV @ 5.9 keV | Available | |
High Resolution X-ray Microscope: - PCO CCD camera 4008 x 2672 pixels, 14bit dynamic range - Objectives :1.25X to 40X - Several optimized scintillators - Theoretical spatial resolution limit <0.5 μm | Available | |
X-ray Eye: Photonic Science X-ray MiniFDI, 11mm dia, 1392 x 1040 pixels, 6.5μm optical pixel size | Available | |
APD Ultrafast X-ray detector Active Area : 10mm x 10mm Max count rate: 100 MHz | Available | |
Other Detectors: PIPS diodes, scintillation counter and ionization chambers | Available | |
Equipment | Availability | Comments |
5-circle Huber Diffractometer with Chi cradle Detector mass ≤30 kg Theta (ω) : -100° to +235.5° 2Theta (δ) : -10.4° to +101.9° Alpha : -33° to +33° Chi : -96.4° to +259.9° Phi : -147° to +150° Theta-analyser : 0 to ∞ (no limits), 2Theta-analyser: -157.5° to +157.5° | Available | - Angular range of some axes is restricted by mounts and beamline delivery systems. Available angular ranges are given here. -2Theta (δ) range can be changed/adjusted, if needed. |
XYZ sample stage on Diffractometer X = ±25 mm Y = ±25 mm Z = ±12.5 mm Load capacity ≤20 kg | Available | |
Versatile Optics Table (with 3 stations) - Size: 2m x 1.2m Upstream station X = ±25 mm Y = ±25 mm Theta = ±135° Phi = ±10° Chi = ±10° Load capacity ≤5 kg Middle station X = ±25 mm Y = ±25 mm Theta = ±135° Phi = ±10° X Piezo=300 µm Y Piezo=300 µm Load capacity ≤5 kg Downstream station X = ±50 mm Y = ±50 mm Load capacity ≤ 20 kg Relative motions (along beam direction and transverse) possible between the three stations. | Available | |
Fast shutter (6mm opening aperture, to be used with focused or slitted beams) | Limited Availability | |
Cryostream Plus (80 – 500 K) - to cool small samples by a cryo-stream | Available | |
Various general-purpose viewing cameras | Available | |
Other Equipment: Huber 5 axis tower Several linear and rotary stages Several Piezo linear stages | Available | Contact B16 Staff. |
Equipment | Availability | Comments |
Zeiss Stereo Optical Microscope | Available | |
Laminar Flow table | Available | |
Fume cupboard, small balance (mg – g), ultrasonic bath, fridge, etc | Available | |
Dark room for photographic film development | Available |
For any further information, please contact a member of the beamline staff.
Diamond Light Source is the UK's national synchrotron science facility, located at the Harwell Science and Innovation Campus in Oxfordshire.