Welcome to I03
8th June 2020: Access to MX beamlines is currently operating under new guidelines. Please read carefully.
I03 is one of five macromolecular crystallography (MX) beamlines at Diamond. I03 is a tuneable beamline with a working wavelength range of 0.6 - 2.48 Å. The standard working wavelength is 0.976 Å (12.7 keV) with a focused beam size of 90 x 30 um (FWHM). As with all the MX beamlines, I03 uses SPINE standard pins and Unipucks for cryo-cooled samples.
I03 can provide containment measures for experiments involving biological agents in Hazard Groups 2 and 3.
Please note there are some adjustments required for Data Processing from i03 Eiger Data from mid 2021.
Automated Single Crystal Crystallography
I03 Specifications
Standard energy (Å / keV) | 0.98 / 12.7 |
Flux at 12.7 keV, focus at sample (ph/s) | 1 x 1013 |
Beam size options:
|
90 x 20 µm 20, 50 & 100 µm |
Wavelength range (Å) | 0.6 - 2.3 |
Energy range(keV) | 5.5 - 25 |
Detector | Eiger2 XE 16M |
Time to collect 180 ° data set (seconds) | 3.2 + |
Sample Changer | BART |
No. of pins in dewar | 592 |
Sample exchange time (seconds) | 18 |
Guide to samples screened/hour (X-ray Centring) | 30 |
+ 1800 frames, 0.1 ° oscillation at 560Hz
Features include:
- Containment for Hazard Group 2 and Hazard Group 3 biological agents
- Apertures (20, 50 or 100
µm) . - BART sample changer - 37 Unipucks / 592 samples
- Eiger2 XE 16M detector - maximum frame rate of 560 Hz
- Vortex fluorescence detector
- Automated data processing
- Possibility to focus at sample position or at detector
- Remote access