Parameter | Availability | Comment |
Spectral range | available | (1-2000 μm) 10000-5 cm-1 1.2 ev - 0.6 mev |
special beam conditions | 5 - 100 cm-1 Coherent emission | |
Beam size at sample | available | FWHM = 15 μm with 36x objective Aperture selectable at sample from 3 - 15 μm |
Beam stability at sample stage | available | <0.1 μm |
Max photon flux at beamline exit normalised to Diamond storage ring at 300 mA | available | In vacuum: 3.8.1013 ph/sec 0.1%bw @ λ=2.1 μm 3.0.1013 ph/sec 0.1%bw @ λ=10.8 μm |
Max photon flux at sample normalised to Diamond storage ring at 300 mA | available | At FTIR in vacuum: 9.2.1012 ph/sec 0.1%bw @ λ=2.1 μm 8.1.1012 ph/sec 0.1%bw @ λ=10.8 μm |
Max photon flux at sample normalised to Diamond storage ring at 300 mA | available | At microscope via 36x objective: 9.0.1011 ph/sec 0.1%bw @ λ=2.1 μm 7.5.1011 ph/sec 0.1%bw @ λ=10.8 μm |
Flux at sample | available | >7.5 x 1011 ph/sec @ 0.1%bw and λ<10 μm. |
Equipment or Component | Availability | Comments |
Bruker Vertex 80 V Fourier Transform IR Interferometer | available | Under vacuum Max resolution 0.07 cm-1 Step scan option Time resolved scan >12 µs |
Beamsplitters | available | UV, visible, near-, mid-, far-IR (5 cm-1) |
Detector | available | Internal DTGS detector for mid- and far-IR regions Broadband MCT near-mid-IR External 4.2 and 1.6 K bolometer for far-IR |
Hyperion 3000 microscope | available | Transmission/reflection IR mode Camera for visible imaging Mapping software View-through 4 blade slits, metal 4-blades transmission/reflection slits |
Objectives/Condensers | available | 4x, 8x, 20x and 40x visible objectives 15x, 20x, 36x and 74x IR objective and condensers Grazing angle 15x IR objective μATR objective in Si and Ge |
Polarisers | available | Visible and IR linear polariser / analyser |
Detectors | available | 64x64 and 40 μm pixel size Focal Plane Array for near- and mid-IR 100x100 μm2 MCT broadband detector for mid-IR (down to 500 cm-1) External 4.2 and 1.6 K bolometer for far-IR |
Sample stage | available | Micrometric high precision x-y and z remotely controlled N2 Purged box |
Equipment or Component | Availability | Comments | |
Bruker Vertex 80 V Fourier Transform IR Interferometer | available | Under vacuum Max resolution 0.07cm-1 Step scan option Time resolved scan > 12µs | |
Beamsplitters | available | UV, visible, near-, mid-, far-IR (5 cm-1) | |
Detector | available | Internal broadband MCT detector for near- and mid-IR regions External 4.2 and 1.6 K bolometer for far-IR* | |
Hyperion 3000 microscope | available | Transmission/reflection IR mode Camera for visible imaging x-y and z remote control mapping software view-through 4 blade slits, metal 4-blades transmission slits | |
Objectives/Condensers | available | 4x, 8x, 20x and 40x visible objectives 15x, 20x, 36x and 74x* IR objectives and condensers Grazing angle 15x objective* μATR objective* in Si and Ge | |
Polarisers | available | Visible and IR linear polariser/ analyser | |
Detectors | available | 64x64, 40µm pixel size Focal Plane Array for near- and mid-IR regions 50x50 μm2 high sensitivity MCT detector for mid-IR region (down to 650 cm-1) External 4.2 and 1.6 K bolometer for far-IR | |
Sample stage | available | Micrometric high precision x-y and z remotely controlled N2 Purged box |
Equipment or Component | Availability | Comments |
Compression cell | available | Suite of windows available |
Cooling/heating sample cell | available | Linkam cell: T from -196 - 600 C |
Diamond compression cell | available | Room temperature |
Liquid flow cell | available | Demountable liquid cells for 13 mm or 25 mm diameter windows. Temperature controlled demountable liquid cell for 13 mm diameter windows. Suite of spacers 6 - 950 µm |
Inverted visible microscope | available | Phase contrast Magnification from 10x - 40x |
Stereo Zoom microscope | available | Reflection and transmission 16 to 1 zoom for large objects Fiber optics illumination |
High resolution microscope | available | 4x to 50x objectives, 100x immersion objective Reflection and transmission upright microscope Bright and dark field, phase contrast, polarisers and DIC High resolution camera Dedicated mapping and analysis software |
available | Fluorescence imaging Filter set 02 (DAPI EX G 365, BS FT 395, EM LP 420) Filter set 09 (EX BP 450-490, BS FT 510, EM LP 515) Filter set 14 (EX BP 510-560, BS FT 580, EM LP 590) | |
Cryo-microtome | available | Diamond, Tungsten Carbide, glass and stainless steel blades LN2 cooled |
Pellet press | available | 10 tonnes max 13 mm and 8 mm pellet |
Chemical fume cupboard | available | For inorganic sample preparation |
BioHazard Cabinet Class II | available | For containment level 1 sample preparation |
Incubator | available | For cell culture containment level 1 and requiring CO2 Incubator with O2 available on demand |
Centrifuges | available | Temperature controlled for Falcon tubes Minispin for Eppendorf vials |
Fridge/freezer | available | Sparkfree Fridge 4 °C Freezer -25 °C |
Freezer -80 °C | available on demand | On demand available as external facility |
Drying Oven | available | To 300 °C |
Analytic and general balances | available | |
Water bath and ultrasound bath | available | |
Hot plate, stirrers | available | |
MilliQ, autoclaved water | available |
Inverted visible microscope | available | Phase contrast Magnification from 5x - 40x Live digital camera |
BioHazard Cabinet Class II | available | For containment level 2 sample preparation |
Incubator | available | For cell culture containment level 2 and requiring only CO2 |
Centrifuges | available | Temperature controlled for Falcon tubes Minispin for Eppendorf vials Cytospin for slide preparation |
Fridge/freezer | available | Sparkfree Fridge 4 °C Freezer -25 °C |
Freezer -80°C | available on demand | On demand available as external facility |
Water bath, water purifier MilliQ | available | |
Automated cell counter | available | |
Cryogenic vessels (40L and 1.5L) | available | For sample storage |
Diamond Light Source is the UK's national synchrotron science facility, located at the Harwell Science and Innovation Campus in Oxfordshire.
Copyright © 2022 Diamond Light Source
Diamond Light Source Ltd
Diamond House
Harwell Science & Innovation Campus
Didcot
Oxfordshire
OX11 0DE
Diamond Light Source® and the Diamond logo are registered trademarks of Diamond Light Source Ltd
Registered in England and Wales at Diamond House, Harwell Science and Innovation Campus, Didcot, Oxfordshire, OX11 0DE, United Kingdom. Company number: 4375679. VAT number: 287 461 957. Economic Operators Registration and Identification (EORI) number: GB287461957003.