Beamlines | I11 - Facility and Equipment availability

Last updated: 2 November 2011

The energy of the beamline is normally set at 15 keV (lam ~ 0.826 Ang) for all experiments.

With the standard beamstop in position the lowest accessible 2 theta angle is ~ 3 degrees (d ~ 15 Ang at 15 keV).

If users would like to operate at different energies, or require access to lower 2 theta angles, please inform your local contact(s) well in advance of the beamtime.

Endstation

Equipment or component
Availability
Comments
Diffractometer (3 circles)
Available
 
High resolution mode: Multi-analysing crystals (MAC): x5 MAC-arms with 45 MAC channels.
Typical collection time per powder pattern: 30 - 60 minutes.
MAC1
Available
Designed 2Θ coverage: 0 - 30°
MAC2
Available
Designed 2Θ coverage: 30 - 60°
MAC3
Available
Designed 2Θ coverage: 60 - 90°
MAC4
Available
Designed 2Θ coverage: 90 - 120°
MAC5
Available
Designed 2Θ coverage: 120 - 150°
Large sample table
(max. load: 500 kg)
Available
 
Small sample table
(max load: 5 kg)
Available
 
Capillary spinner
Available
 
Flat-plate spinner
Available
 
Time-resolved mode:
PSD (Mythen2)
Typical collection time per powder pattern: ms - minutes.
Available
Angular coverage: 2Θ=90°

Sample environments and stages

Robotic arm & carousel
(max. 200 capillary samples)
Available
Cryostream Plus (80 – 500 K)
Available
Hotair blower (RT – 1000 °C)
Available
Capillary furnace
(RT-1500 °C)
Available
PheniX cryostat (11 – 295 K)*
Available
Flat-plate furnace, MRI
(295 -1700 °C)
Available
DSC (Linkam)
Available
5K Cryostat (5 – 295 K)**
Available
Humidity chamber
Available
Low Pressure Capillary Gas Cell (mbar – 1.5 bar)  Available
High Pressure Capillary Gas Cell (mbar – 100bar)  Available

* Please allow at least three hours to change samples.

**Allow one shift of beamtime to allow for set-up and cool-down. Maximum sample turnover one per day.

Support Lab

Basic chemistry lab.
Available
With basic equipment and common solvents
Kit for loading capillary and flat-plate samples
Available
Pestle, mortar, tweezers, spatulas, etc.
Fume cupboard
Available
 
De-ionised water
Available
 
Ultra-sonic baths
Available
 
Small balances (mg – g)
Available
 
Fridge & Freezer
Available
 
Glove Box Available by prior arrangement If the use of a glove box is required please contact beamline staff in advance of your experiment with further details so that this may be arranged.

 

Silicon pattern (SRM 640c): collection time = 30 mins, l=0.826982 ÅSilicon pattern (SRM 640c): collection time = 30 mins, λ=0.826982 Å

I11 Beamline Manual